Operation
51
Manual ranging requires the range to be selected by the
user. For more information, refer to “Up and Down
Arrows” on page 40.
Zeroing
Thermopile and optical sensors require periodic zeroing.
Zeroing occurs when a meter attempts to null out any
signal coming from the sensor and establish a zero-
power baseline. If Auto Ranging is active, the sensor is
zeroed for every available range. If Auto Ranging is not
active, the sensor is zeroed at the current range only.
Under typical operating conditions, the zero procedure
takes about one second for each range.
When starting the zeroing procedure, the large numerals
on the LCD are replaced by an animated set of dashes.
Unless there is an error, the normal measurement mode
resumes once the zeroing procedure ends.
Figure 16. Auto Ranging